4 ADDITIONAL ADVICE FOR OPERATION
The nger should be put properly and correctly.
Avoid shaking nger as possible as you can during measurement;
Do not put wet nger directly into sensor.
Avoid placing the device on the same limb which is wrapped with a cuff for
blood pressure measurement or during venous infusion.
Do not let anything block the emitting light from device.
Electrosurgical device interference may affect the measuring accuracy.
Using enamel or other makeup on the nail may affect the measuring accura-
cy.
If the rst reading appears with poor waveform (irregular or not smooth), then
the reading is unlikely true, the more stable value is expected by waiting for a
while, or a restart is needed when necessary.
5 TECHNICAL SPECIFICATIONS
A. Display mode: Color OLED Display
B. Power supply requirement:
2 x LR03 (AAA) alkaline batteries or Ni-MH rechargeable batteries
Supply voltage: 3.0 VDC
Continues working time (display is off): >12hours
C. Operating current: ≤50mA
D. SpO2 Parameter Specications
Transducer: dual-wavelength LED
Measurement wavelength:
Red light: 663 nm, Infrared light: 890 nm.
Maximal optical output power: less than 1.5mW maximum average
Measuring range: 35~100%
Measuring accuracy:
Not greater than 3% for SpO2 range from 70% to 100%
*NOTE: Arms is dened as root-mean-square value of deviation according to
ISO 9919.
E. Pulse Rate Parameter Specications
Measuring range: 30bpm~240bpm
Accuracy: ±2bpm or ±2% (whichever is greater)
F. Preset limits:
SpO2 Low limit setting: 85%
Pulse Rate limit setting range:
Upper limit: 120bpm
Lower limit: 50bpm