54
Determined by the diagnostic result that judged as normal in cardiac
catheterization and ultrasonic examination, and the result that judged as
normal in physical examination.
2) Atrium hypertrophy
Determined by the diagnostic results of ultrasonic examination.
3) Myocardial infarction and myocardial ischemia
Determined by the physician diagnostic results of cardiac catheterization.
4) Tachycardia, bradycardia, low voltage, axis
Determined by the diagnostic results of ultrasonic examination.
5)Conduction block
Determined by the physician diagnostic results of cardiac catheterization.
The standard of normal population in the customized database: physical
examination is normal, no heart disease or other diseases that may affect
cardiac functions or shape.
4.5 Data coverage of verification for automated interpretation
Analyzing the content of CSE diagnostic database and customized data, the overall
condition and coverage of statistical samples are shown as below: