IMMUNITY TEST IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT
GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, con-
crete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
N/A
N/A
Surge
IEC 61000-4-5
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
N/A
N/A
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0,5 cycle
<40% U
T
(60% dip in U
T
)
for 5 cycles
<70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
N/A
N/A
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
elds should be at levels cha-
racteristic of a typical location
in a typical commercial or ho-
spital environment.
Guidance and manufacturer’s declaration – electromagnetic emission
The Fingertip Oximeter is intended for use in the electromagnetic environments
specied below, the customer or the user of the Fingertip Oximeter should
assure that it is used in such an environment.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.