113
IMMUNITY TEST IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT
GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, con-
crete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
N/A
N/A
Surge
IEC 61000-4-5
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
N/A
N/A
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0,5 cycle
<40% U
T
(60% dip in U
T
)
for 5 cycles
<70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
N/A
N/A
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
elds should be at levels cha-
racteristic of a typical location
in a typical commercial or ho-
spital environment.
Guidance and manufacturer’s declaration – electromagnetic emission
The Fingertip Oximeter is intended for use in the electromagnetic environments
specied below, the customer or the user of the Fingertip Oximeter should
assure that it is used in such an environment.
Note:
U
T