3. TECHNICAL SPECIFICATIONS
A. Display mode: OLED Display
B. Power supply requirement:
1.5V (AAA size) alkaline battery x 2
Battery voltage: 3VDC
C. Operating current: <50mA
D. SpO2 Parameter Specifications:
Measuring range: 35%~100%
Accuracy: ≤3% (during 70% - 100%)
E. Pulse Rate Parameter Specifications:
Measuring range: 30bpm - 240bpm
Accuracy: ±2bpm or ±2% (which ever is greater)
Pulse Rate alarm: Upper limit: 120bpm
Lower limit: 50bpm
F. Blood Perfusion Parameter Specifications:
Measuring range: 0%~20%
G. The performance under low perfusion condition
The accuracy of
SpO2
and PR measurement still meets the specica-
tion described above when the modulation amplitude is as low as 0.6%.
H. Resistance to ambient light interference:
The accuracy of
SpO2
and PR measurement still meets the speci-
cation described above when the device is tested by
SpO2
simulator
(Fluke Biomedical Index 2 series) while setting the emulating interfer-
ence of sun light and 50Hz/60Hz uorescent light.
I. Dimensions: 66 mm (L) x 36 mm (W) x 33 mm (H)
Net Weight: 60g (including batteries)
J. Classification:
The type of protection against electric shock: Internally powered
equipment.
The degree of protection against electric shock: Type BF equipment.
The degree of protection against harmful ingress of water: Ordinary
equipment without protection against ingress of water.
Electro-Magnetic Compatibility: Group I, Class B