Maximal R-R Limit: set a percentage value, the ECG waveform with RR interval more than
average RR interval by this percentage value will be rejected.
Minimal R-R Limit: set a percentage value, the ECG waveform with RR interval less than
average RR interval by this percentage value will be rejected.
Note: Auto check function is useable in the interface of wave affirm.
Wave affirm
Waves can be filtrated by "auto check" or manual (as Fig.6-58).
RR interval: display RR interval trend graph of all accepted beats in waveform window.
Waveform: display 3 leads waves after file conversation. Herein the white curves denote
accepted waves, and the grey waves denote rejected waves. Click the left key of mouse on
waveform to change the selection state of into accepted or rejected state beat on this position.