EasyManuals Logo

Gima TLC5000 User Manual

Gima TLC5000
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
1
Appendix IV Accuracy of Operating Data
The systems having automated ECG analysis is required to fulfill of this table
Gross
statistics
AHA DB MIT DB NST DB CU DB
The
assess
ment
criteria
require
d (%)
The
actual
test
results
(%)
The
assess
ment
criteria
require
d (%)
The
actual
test
results
(%)
The
assess
ment
criteri
a
requir
ed (%)
The
actual
test
result
(%)
The
assess
ment
criteria
require
d (%)
The
actual
test
result
s (%)
QRS Se (G) 98 98.91 98 98.79 83 83.44 -- --
QRS +P (G) 98 98.70 99 99.70 94 94.57 -- --
VEB Se (G) 83 83.45 83 83.82 79 79.82 -- --
VEB +P (G) 89 89.86 84 84.31 68 68.14 -- --
SVEB Se (G) -- -- 18 18.90 37 37.33 -- --
SVEB +P (G) -- -- 10 10.37 29 29.94 -- --