5.4 The Problem Related to the Cable and the Input Plug
Problem
Cause
Correction
The output wave of the
recorder is a straight
line.
1. The recorder isn't connected
well.
Please check if the needles of
the plug is curved, broken or
lack. If the plug is well, please
connect once again.
2. The cable is broken. Please contact our company.
3. The recorder is broken. Please contact our company.
Some ECG wave is
disturbed a lot, and the
quality of ECG signal
is poor.
1. The cable isn't connected
well
Connect the cable once again
according to the operation
manual.
2. Lead cable is broken. Please change a new one.
3. The quality of the one-off
electrode is poor.
Please change new,
high-quality electrode.