9
2. Press M to mark this test as a control solution test
Press M until “QC is displayed on the screen, to ensure that the test result
is not stored in the meter memory. If you press M again, the “QC will
disappear and this test is no longer a control solution test.
WARNING:
When performing a control solution test, you have to mark it accordingly
to ensure that the test result is not stored in the meter memory. Failure to
do so will result in the control solution test result being included with your
blood glucose test results in memory.
3. Apply control solution
Shake the control solution bottle thoroughly before use. Squeeze out the
rst drop and wipe it o, then squeeze out another drop and place it on the
tip of the bottle cap. Hold the meter to move the absorbent hole of the test
strip to touch the drop. Once the conrmation window lls completely, the
meter will start counting down.
NOTE:
To avoid contaminating the control solution, do not directly apply control
solution onto a strip.
4. Read and compare the result
After counting down to 0, the control solution test result will appear on the
display. Compare this result with the range printed on the test strip vial – it
should fall within this range. If not, please read the instructions again and
repeat the control solution test.