log
samp l time
0.35
P
Sampling time
Sec: 0.3, 0.6, 1, 2,
5,10, 15, 30
Min
: 1, 2, 5, 10, 15, 30
0.35
30 m
log
stor e time
1m
P
Wh
Storage rate
Min: 1, 2, 5,10, 15, 30
hour:1, 2, 4, 8, 12
day: 1, 2, 4
1m
4d
Sampling time T
sa
, storage time T
st
and number of
traces Tr result in a maximum number of storable trig-
ger events N with a memory capacity of 512 kByte
(Round N up to whole number: Nmin = 1, Nmax = 99)
If the display blinks when the value is selected,
the memory is too small for the selected setting.
N = (250 000 x T
sa
) / (T
st
x Tr)
GMC-I Messtechnik GmbH 33