46 GMC-I Messtechnik GmbH
Err14
SETUP
Resistance compensation for the connector cables:
R
OFFSET
> 1 :
OFFSET measurement of RL-PE or RN-PE and RLN for ZL-PE and ZL-N is not
sensible.
Remedy: check system.
Err15 R
LO
R
OFFSET
> 10 :
OFFSET measurement is not sensible.
Remedy: check system.
Err16
SETUP OFFSET
(EXTRA U)
Z > 10 :
OFFSET measurement of RL-PE or RN-PE and RLN for U(ZLN) is not sensible.
Remedy: check system.
Err17 EXTRA U
U
OFFSET
> U:
OFFSET value is greater than the measured value at the consuming system.
OFFSET measurement is not sensible.
Remedy: check system.
Err18
RINS
/ RLO
Contact problem or blown fuse
Remedy: check test plug or measuring adapter for correct seating in the test plug,
or replace the fuse.
Err19
R
E
Polarity of the test probes has to be reversed.
Err20 I
N
/ I
F
N and PE are swapped.
Err21
I
N
/ I
F
Z
L-N
/ Z
L-PE
/ R
E
1) Mains connection error
Remedy: inspect mains connection.
or
2) Display in the connection pictograph: PE interrupted (x) or bottom protective
conductor tab interrupted with reference to the keys at the test plug
Cause: voltage measuring path interrupted
Result: measurement is disabled.
Note: only if appears: Measurement can nevertheless be started by pressing
the ON/START key again.
Err22
I
N
/ I
F
Display in the connection pictograph:
Top protective conductor tab interrupted with reference to the keys at the test plug
Cause: current measuring path interrupted
Result: no measured value display
Err23 I
N
/ I
F
Resistance in the N-PE path is too high.
Consequence: The required test current cannot be generated and measurement is
aborted.
Err24 Z
L-PE
, R
E
If specified touch voltage U
L
is exceeded:
Z
L-PE
and R
E
: user is prompted to switch to the 15 mA wave.
R
E
alternative only:
User is prompted to reduce the measuring range (reduce current.)
Status
Error
No.
Position of the
Function Switch
Function / Meaning
I
N
/I
F
10 mA 30 mA 100 mA 300 mA 500 mA
R
MAX
at I
N
510 170 50 15 9
R
MAX
for I
F
410 140 40 12 7