APC-US-INS-LAB-RevC18 34
9.7 RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND
MOBILE RF COMMUNICATIONS EQUIPMENT AND HAUSTED APC SERIES
MODELS AND SURGI-CHAIRS
The Hausted APC Series Models and Surgi-Chairs are intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the Hausted APC Series Models
and Surgi-Chairs can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Hausted APC Series Models
and Surgi-Chairs as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.7 GHz
d = 2.3√P
0,01 0.12 0.12 0.23
0,1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reection from structures, objects, and people.
9.8 NOTES TO SECTION 9
1. 60601-1-2 © IEC:2014

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Graham Field HAUSTED EPC Series and is the answer not in the manual?

Graham Field HAUSTED EPC Series Specifications

General IconGeneral
BrandGraham Field
ModelHAUSTED EPC Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals