23. ERROR MESSAGES
ADC fault: No response from
ADC.
Malfunction in scale. Consult
helpdesk.
Sensor input signal too high
(>20mV).
Faulty load cell. Cabling in
poor condition.
Sensor input signal too low
(<-20mV).
Faulty load cell. Cabling in
poor condition.
Dead battery.
Connect the AC/DC adapter
to charge the battery.
a) Auto-zero device out of
range.
b) A stable measurement
could not be obtained when
attempting initial zeroing.
Empty the load receptor
before switching on the scale.
Calibration weight > Max.
Calibration weight should be
< Max.
The division should be >10 ADC
counts.
Resolution is too high.
Change the division to a
higher value.
A stable measurement could
not be obtained for setting the
scale.
Overload: Weight > Max+9·div
The keypad locking function is
activated.
Unlock the keypad by keeping
the key pressed for two
seconds.
The optional RTC is not present
or the circuit is damaged.
Consult helpdesk.
A totalizer ticket has now been
initiated.
Print the total before
initiating a new ticket.