JB Academy, JB Nova and SUB Aqua Pro 31394 V7
Detailed operating manual Page 1
www.grantinstruments.com
Unstirred Water Baths
JB Academy
JB Nova
SUB Aqua Pro
Operating Manual
www.grantinstruments.com
salesdesk@grantinstruments.com
Tel: +44 (0)1763 260811

Table of Contents

Questions and Answers

  • A
    Alexander HernandezAug 9, 2025
    Why temperature continues to rise when not expected in Grant Laboratory Equipment?
    • J
      Jason GrantAug 9, 2025
      If the temperature continues to rise unexpectedly in your Grant Laboratory Equipment, it may be because the set temperature is higher than the liquid temperature. Ensure the bath set temperature is correct. Alternatively, the set temperature might be too close to the ambient temperature. In this case, increase the set temperature, keeping in mind that the minimum temperature should be 5°C above ambient.
  • W
    William JonesAug 17, 2025
    How to resolve if temperature does not rise when expected in Grant Laboratory Equipment?
    • M
      Melvin DennisAug 17, 2025
      If the temperature in your Grant Laboratory Equipment isn't rising as expected, the set temperature might be lower than the liquid temperature. Verify that the bath set temperature is correct. Also, if the set temperature is too close to ambient, increase it, noting that the minimum temperature should be 5°C above ambient.
  • S
    sanderswilliamAug 25, 2025
    What to do if Grant Laboratory Equipment display shows '????'?
    • J
      Joshua OlsenAug 26, 2025
      If the Grant Laboratory Equipment display shows '????', it could indicate a faulty temperature probe. A competent person should check the probe for an open or short circuit fault, or contact Grant.
  • J
    joshua48Sep 1, 2025
    What to do if the Grant Laboratory Equipment display shows '???'?
    • A
      amanda56Sep 1, 2025
      If the Grant Laboratory Equipment display shows '???', it means that the bath is running on its default settings. Contact Grant Instruments' Service Department for assistance.

Summary

Safety Information and Warnings

4.1 Safety Compliance

Details compliance with international safety standards like IEC 61010 for measurement equipment.

4.3 Essential Safety Warnings

Provides critical warnings and precautions for safe operation and handling of the water bath.

Operating Procedures and Modes

6.1 General Operation

Covers water level, dry start protection, operation above 60°C, and flat bottomed vessels.

Model-Specific Operation

7.3 SUB Aqua Pro Operation

Covers controls, temperature settings, keypad lock, over temp protection, presets, and timer for SUB Aqua Pro.

Support and Compliance

12 Troubleshooting Guide

Provides solutions for common symptoms like temperature issues and display errors.

Grant JB Nova Series Specifications

General IconGeneral
BrandGrant
ModelJB Nova Series
CategoryLaboratory Equipment
LanguageEnglish