EasyManuals Logo

Grant JB Nova Series User Manual

Default Icon
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
JB Academy, JB Nova and SUB Aqua Pro 31394 V7
Detailed operating manual Page 20
www.grantinstruments.com
8.3 Dual point calibration (, )
Dual point calibration is typically used on baths which are operated over a range of
temperatures.
Two calibration points are entered into the calibration menu: the low calibration point
() and high calibration point (). These points are typically selected as just below
and above the normal working temperature range for the bath.
Note that if only one calibration point is entered then the bath will behave as described in
Single point calibration above.
To configure a dual point calibration, follow the steps below.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Grant JB Nova Series and is the answer not in the manual?

Grant JB Nova Series Specifications

General IconGeneral
BrandGrant
ModelJB Nova Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals