8 Calibration
8.1 Calibration options
Two calibration options exist:
Single point calibration (available on all models)
Dual point calibration (available on SUB Aqua Pro models only)
The quality of the calibration is highly dependent on:
Use of a suitable reference thermometer, ideally 10 times the accuracy
you are trying to achieve.
Performing a calibration in a stable ambient environment (+/-1ºC) free
from draughts or cooling air currents.
8.2 Single point calibration (and )
A single point calibration applies a single offset over the bath temperature curve. For this
reason the calibration temperature is usually the same as the intended working
temperature for the bath or particular experiment: