JB Academy, JB Nova and SUB Aqua Pro 31394 V7
Detailed operating manual Page 1
www.grantinstruments.com
Unstirred Water Baths
JB Academy
JB Nova
SUB Aqua Pro
Operating Manual
www.grantinstruments.com
salesdesk@grantinstruments.com
Tel: +44 (0)1763 260811

Table of Contents

Other manuals for Grant SAP34

Questions and Answers

  • L
    Lee HollandAug 10, 2025
    Why temperature continues to rise when not expected in Grant Laboratory Equipment?
    • M
      Michael CrossAug 10, 2025
      If the temperature continues to rise unexpectedly in your Grant Laboratory Equipment, it may be because the set temperature is higher than the liquid temperature. Ensure the bath set temperature is correct. Alternatively, the set temperature might be too close to the ambient temperature. In this case, increase the set temperature, keeping in mind that the minimum temperature should be 5°C above ambient.
  • B
    Brian JohnsonAug 17, 2025
    How to resolve if temperature does not rise when expected in Grant Laboratory Equipment?
    • E
      Eric AlvarezAug 17, 2025
      If the temperature in your Grant Laboratory Equipment isn't rising as expected, the set temperature might be lower than the liquid temperature. Verify that the bath set temperature is correct. Also, if the set temperature is too close to ambient, increase it, noting that the minimum temperature should be 5°C above ambient.
  • C
    chelseaharrisAug 25, 2025
    What to do if Grant Laboratory Equipment display shows '????'?
    • C
      Charlotte BakerAug 25, 2025
      If the Grant Laboratory Equipment display shows '????', it could indicate a faulty temperature probe. A competent person should check the probe for an open or short circuit fault, or contact Grant.
  • M
    Michael NicholsonSep 1, 2025
    What to do if the Grant Laboratory Equipment display shows '???'?
    • A
      anthony59Sep 1, 2025
      If the Grant Laboratory Equipment display shows '???', it means that the bath is running on its default settings. Contact Grant Instruments' Service Department for assistance.

Summary

Safety Information

Safety Symbols and Warnings

4.3 Safety Warnings

Provides critical safety warnings for operation, handling, and environment.

Operating Instructions

5.4 Electrical Supply

Checks for correct supply voltage, plug type, and grounding connection.

Operating Procedures

6.1.2 Operating the Bath Without Water (Dry, DEC)

Warns against dry operation and explains the dry start protection mechanism.

6.1.5 Allowing the Bath to Run Dry (DPA)

Explains the advanced dry protection alarm and how to disable it.

Using the Water Bath

7.1 Using the JB Academy

Details operation of the JB Academy water bath.

7.1.3 Setting the Control Temperature (°C)

Step-by-step guide to setting the desired water temperature for the JB Academy.

7.2 Using the JB Nova

Details operation of the JB Nova water bath.

7.2.2 Setting the Control Temperature (°C) and Keypad Lock

Guide to setting temperature and locking the keypad on the JB Nova.

7.3 Using the SUB Aqua Pro

Details operation of the SUB Aqua Pro water bath.

7.3.2 Setting the Control Temperature (°C)

Guide to setting the desired water temperature for the SUB Aqua Pro.

7.3.4 Setting the Over Temperature Protection (DEP)

Instructions for setting a maximum temperature limit to protect samples.

Calibration

8.1 Calibration Options

Outlines the two available calibration methods: single point and dual point.

8.2 Single Point Calibration (SPC and LPC)

Details the process and considerations for single point calibration.

8.3 Dual Point Calibration (LPC, HPC)

Details the process for dual point calibration used for wider temperature ranges.

Technical Specifications

Maintenance and Service

Troubleshooting