CPT20 USER MANUAL
8
When the scale is brand new and has not been calibrated; it will start automatically
with Spec Calibration and Weight Calibration.
In weighing mode:
Press and hold T key and 1 key Spec and Weight Calibrations
(CAL_ON Calibration switch at ON position)
Press and hold T key and 2 key Weight Calibration
(CAL_ON Calibration switch at ON position)
Press and hold T key and 3 key in weighing mode G value look up or
calibration. Spec and Weight Calibrations (CAL_ON Calibration switch at ON
position or G value calibration set as non-approval standard calibration)
Press and hold T key and 4 key Standard Functions Setup
Press and hold T key and 5 key Keyboard test, displays calibration weight internal
resolution and version number
Press and hold T key and 6 key Displays battery voltage, internal value and ADC
internal value
Spec and Weight Calibrations
(CAL_ON Calibration switch at ON position)?
Key description
Enter
0-9
Numeric Keys
Left
Right
Escape
The first two digits are for division setup.
The third digit is for unit price and total price decimal point (cent) setup.
The fourth digit is for G value On or Off for external calibration
The fifth digit is for P1/P2 setup
Zero
C E