RS485 SERIAL OUTPUT
(Replaces standard RS232 Output with Data Logger option.)
Permits transfer of optional Data Logger reports on wire lengths up to 4000 ft. (1220 m).
Format: 8 Bits, 1 Stop Bit, No Parity. (Baud rate may be changed under the
SPECIAL FUNCTIONS
menu. Factory default is 19200 baud).
Use shielded, 4-conductor cable (Greyline cable option SC-18AWG recommended). Connect the cable
shield to Ground at the DFM-IV electronics enclosure and do not connect to Ground at the other end.
Configure the Greyline software program for RS485 communications. Under the Communications /
Connection Setup menu, select 'Connected to a Greyline 485OI-12V adaptor'.
4850I9TB-12V
RS232/RS485 Converter
Use with optional RS485 output for connection to a PC
computer or modem.
•
Optically isolates and protects your computer's RS232
port
•
Terminal block for RS485 connections
•
Operates from 2400 up to19.2K baud
DB9M/DB9F - RS232 Cable
Connects RS232/485 Converter to a PC
Cable length 6 ft. (2 m) with DB9M and DB9F connectors.
DFM-IV Doppler Flow Meter
Manual Series A.6
Page 38
Greyline
Transmitter
with
RS485 Output
RS485 <4000 ft (1220 m)
Detail
RX
RS485
PU
GND
TX
(GREEN)
(WHT)
(BLK)
(RED)
18AWG, 4-conductor, shielded
Option: SC-18AWG
DB9F
DB9M
RS232 Cable
Option: DB9M/DB9F
6ft(2m)
RS232/485 Converter
Option: 485OI9TB-12V
RX
TX
GND
PU
RS485
RS232
DB9F