RELAY PARAMETERS
Set Relays 1-3 (4-5 optional) to Off , Pulse or Flow.
P
ulse
press È and set digits to the flow volume increment required between relay
pulses. Use this feature for remote samplers, chlorinators or totalizers.
Minimum time between pulses is 1 second and pulse duration is 350
milliseconds.
Flow
press È and set digits to the required On and Off set points. Use this feature for
flow control and alarms.
Page 9
DFM-IV Doppler Flow Meter
Manual Series A.6
RELAY PARAMETERS RELAY PARAMETERS
Pulse
FlowR1 Function Off
R1on 100 ft3
R1on 100.0 ft3/s
R1 0.00OFF ft3/s
R2
R3
R4
R5
Store? Yes
*** STORING ***