DG Spin
Installation
5-3
• Wait for 30 seconds and verify that the
DG Spin
Head has stopped and
the Display again shows the pre-established centrifugation time.
The
DG Spin
centrifuge is supplied with a conformity certificate stating that the
centrifugation speed (990 rpm) has been verified using instruments calibrated by
accredited entities. The
DG Spin
centrifuge does not need to be readjusted during
its useful life if it is used under normal working conditions.
5.5 Configuration of the equipment
The
DG Spin
centrifuge has some predefined parameters that may be modified by
the OPERATOR. Predefined values are the following:
- Acoustic signal volume: High
- Centrifugation speed units: rpm
- Date and time: GMT time zone.
NOTE:
The Date and Time parameters are used to
maintain traceability in a "log file" with relevant
information and possible incidences that have
occurred during the centrifugation phase.
Accessibility to this file and its use are for the
exclusive use of the authorised Technical
Service.
To modify these parameters access the Configuration mode as described below:
• With the equipment switched off, keep the
button pressed down
and switch on the Main switch (Ref. (6), Figure 2).
The
DG Spin
centrifuge is initialised
in
Configuration mode and
enables personalising the parameters described:
• In Configuration mode the Display will show the configuration
menu for the selection of the acoustic signal volume.
The numeration shown corresponds to the following volumes:
- VOL 0: Deactivated
- VOL 1: Medium level volume
- VOL 2: High volume