6.02 Method of cleaning and disinfection
Cleaning
We recommend that any parts of the mobile lifter that users/helper may come
in contact with are cleaned with Novadan Des Wipes, Novadan Disinfect O or
similar disinfection product.
Novadan Des Wipes:
Use a disposable wipe for cleaning and disinfection of all types of surfaces,
machines and instruments in the healthcare sector.
NB: Do never clean the lifter with strong acid, base or alcohol. Never clean
the lifter in an autoclave machine.
7.00 Troubleshooting
Error:
Lifter does not react to input from the buttons on the hand control.
1. Does the green or yellow LED illuminate on the lifter when the control
buttons on the hand control are activated?
Yes see point 6.
No no light visible – see point 2.
No only red light visible – see point 3.
No yellow service light visible or audible “Beep”– see “indicator and audio
signals matrix”
2. Is the battery installed in the lifter?
Yes see point 4.
No install a fully charged battery.
3. Is a battery with sufcient charge installed in the lifter?
No install a fully charged battery.
4. Is the emergency stop activated?
Yes turn emergency stop in the direction of the arrow to allow it to release
and press any button on the hand control.
No see point 5
5. Is the connector for the hand control tted in the lifter, and is the lifter
reacting to the buttons located on the control box?
Yes replace hand control
No t connector