4.03 Charging procedure
Recharge lifter every night or when it is not in use. This maintains the batte-
ries and ensures a long life span.
Only use power supply provided by Guldmann.
Recharging
• Recharge by plugging the power supply lead
into the wall socket
• Recharging should be carried out when the
yellow or red LED is illuminated
• When the yellow LED illuminates there are
approx. 10 lifts remaining
• When the red LED illuminates and the audio
warning sounds, it is only possible to lower the
patient
• The power supply will turn off automatically
when lifter is fully recharged
• Recharging must not take place in wet rooms/
bathrooms
• Maximum recharging time is approx. 5 hours.
Alternative charging procedure
Remove the battery and place it on a dry surface. Connect power supply plug
to mains.
Now insert the male connector into battery.
Note:
Please be sure that main power is switched off during connection of power
supply and battery.