4.03 Charging procedure
Recharge lifter every night or when it is not in use. This maintains the batte-
ries and ensures a long life span.
Only use power supply provided by Guldmann.
Recharging
Recharge by plugging the power supply lead
into the wall socket
Recharging should be carried out when the
yellow or red LED is illuminated
When the yellow LED illuminates there are
approx. 10 lifts remaining
When the red LED illuminates and the audio
warning sounds, it is only possible to lower the
patient
The power supply will turn off automatically
when lifter is fully recharged
Recharging must not take place in wet rooms/
bathrooms
Maximum recharging time is approx. 5 hours.
Alternative charging procedure
Remove the battery and place it on a dry surface. Connect power supply plug
to mains.
Now insert the male connector into battery.
Note:
Please be sure that main power is switched off during connection of power
supply and battery.