GDM-8245 User Manual
52
Crest Factor Range 3.0 at full scale.
The burden voltage is the same as the DC current.
When the input exceeds the full scale of the selected range, the display will
indicate over-range: “—OL—“.
FREQUENCY MEASUREMENT AT ACA RANGE
RANGE FREQUENCY INPUT LEVEL
(SINE WAVE)
ACCURACY
500uA 10Hz ~ 20kHz
90μA
0.05%+1
5mA 10Hz ~ 20kHz
0.9mA
0.05%+1
50mA 10Hz ~ 20kHz
9mA
0.05%+1
500mA 10Hz ~ 20kHz
90mA
0.05%+1
2A 10Hz ~ 2kHz
1A
0.05%+1
20A 10Hz ~ 2kHz
9A
0.05%+1
AC+DC measurement does not support AC+Hz function.
Resistance
RANGE RESOLUTION ACCURACY
500 0.01
0.1%+4
5k 0.1
0.1%+2
50k 1
0.1%+2
500k 10
0.1%+2
5M 100
0.2%+2
20M 1k
0.3%+2
Open-circuit Voltage
3.2 volts maximum on 500, 1.3 volts maximum on all
other ranges.
Protection 450V dc or peak ac continuous.
Capacitance
RANGE RESOLUTION ACCURACY
5nF * 0.001nF
1nF: 2%+10
<1nF & 0.5nF: 2%+20
50nF 0.01nF
10nF: 2%+10
<10nF & 5nF: 2%+30
500nF 0.1nF 2%+4
5uF 1nF 2%+4
50uF 10nF 2%+4
*5nF range is affected by the impedance of the test lead. For accuracy, please
measure the range directly on the input terminal.
Protection 450V dc or peak ac continuous.