GDS-806/810/820/840 Programming Manual 
 
73
STATus
:QUEStionable :OPERation
Figure 5-2: STATus hierarchy of SCPI defined register
:CONDition
:ENABle
EVENt?
:CONDition :ENABle
EVENt?
:QUEue
[EVENt]?
Enable
Register
Event
Register
Condition
Register
To SBR
 
Figure 5-3: Status registers and related commands 
The CONDition register is a read-only register which monitors the present state of 
the instrument. The CONDition register updates in real time and the inputs are not 
latched or buffered. When a condition monitored by the CONDition register 
becomes true, the bit for that condition also becomes true (1). When the condition 
is false, the bit is 0. 
The read-only EVENt register latches any false-to-true change in condition. Once 
the bit in the EVENt register is set, it is no longer affected by changes in the 
corresponding bit of the CONDition register. The bit remains set until the controller 
reads it. The command *CLS (Clear Status) clears the EVENt registers.