(5) 
Remote control of output current 
The output current of the power supply can be remote
-
controlled with an external voltage, the connection scheme as follow, 
See 
Fig. 
'5
-
5. 
Fig. 
5
-
5 
REAR  PANEL 
REMOTE  CONTROL  VOLTAGE 
SOURCE 
POWER 
A. 
Set the power supply INT
-
SLAVE switch to 
"
PAR
-
SLAVE
" 
position. 
B. 
The 
"-" 
line of control voltage source is connected to the 
"
PAR
n 
input terminal and the 
"+" 
line of control voltage source 
is 
connected 
to 
the power supply output 
" 
+ 
" 
terminal. 
C. 
The 
output current lo calculate 
as 
follows formula: 
lo 
= 
hLEL 
10 
lo:  Output current of the power supply. 
Ir:  Rated current of the power supply. 
Ec:  Remote control voltage. 
OsEcGapprox. 10V 
D. 
For the remote control voltage source Ec,  use a device which provides a stable, low noise voltage source (a 
GOOD 
WILL 
GPR, GPC 
series power supply or an equivalent device). 