GW Instek GPS-3030 - Maintenance and Adjustments; Fuse Replacement; Line Voltage Conversion; Adjustments and Calibration

GW Instek GPS-3030
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
(5)
Remote control of output current
The output current of the power supply can be remote
-
controlled with an external voltage, the connection scheme as follow,
See
Fig.
'5
-
5.
Fig.
5
-
5
REAR PANEL
REMOTE CONTROL VOLTAGE
SOURCE
POWER
A.
Set the power supply INT
-
SLAVE switch to
"
PAR
-
SLAVE
"
position.
B.
The
"-"
line of control voltage source is connected to the
"
PAR
n
input terminal and the
"+"
line of control voltage source
is
connected
to
the power supply output
"
+
"
terminal.
C.
The
output current lo calculate
as
follows formula:
lo
=
hLEL
10
lo: Output current of the power supply.
Ir: Rated current of the power supply.
Ec: Remote control voltage.
OsEcGapprox. 10V
D.
For the remote control voltage source Ec, use a device which provides a stable, low noise voltage source (a
GOOD
WILL
GPR, GPC
series power supply or an equivalent device).