LCR-6000 Series User Manual
67
Soft-key
Function
OFF
Turns OFF the COMP function
ON
Turns ON the COMP function
5.3.2 Auxiliary Bin [AUX]
After AUX is turned on, DUTs that do not fall within
the primary parameter limit values are sorted as OUT. In
addition, DUTs that fall within the primary parameter
limits but out of the secondary parameter limits are
sorted into the auxiliary (AUX) bin.
Procedure for turning ON/OFF the auxiliary bin [AUX]
Step 1. Press the [Setup] key.
Step 2. Press the [BIN MEAS] soft-key.
Step 3. Use the cursor key to select [AUX] field.
Step 4. Use the soft-keys to turn on/off the auxiliary
bin.
Soft-key
Function
ON
Turns ON the auxiliary bin
OFF
Turns OFF the auxiliary bin
5.4 [BIN COUNT] Page
When the [Setup] key and the [BIN COUNT] soft-key is
pressed, the [BIN COUNT] page appears.
This page provides the following information: