GW Instek LCR-914 Series - Short Circuit Calibration

GW Instek LCR-914 Series
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LCR-914/915/916 User Manual
17
WhenSrt” appears on the sub display, short the terminals or the
SMD test probe, and then press the CAL button to start the short
calibration. About 30 seconds later, the result of the short
calibration appears on the main display. If the result is pass,
press the CAL button to complete the calibration. If the result is
fail, press the CAL button to exit the function.