LCR-914/915/916 User Manual
22
Shock Vibration: Sinusoidal vibration per MIL-T- 28800E
(5 ~ 55 Hz, 3g maximum).
Drop Protection: 4 feet drop to hardwood on concrete floor.
Indoor Use.
Electrical Specifications
(1) Test Frequency
[1] The 10kHz test frequency is only for the LCR-915/916.
[2] The 100kHz test frequency is only for the LCR-916.
Accuracy: ±(A x B)(% of reading)
A: Basic Accuracy as specified by
B: Test Cable Accuracy
B(%) = 1 + (L x F x T)
L(m): Cable length
F(MHz): Test frequency
T: Cable type. If the cable is 5-wire type, then “T” is 40, else “T”
is 4,000.
When measuring by basic accuracy, the following
conditions must be met:
1. Ambient temperature: 23˚C ± 5˚C < 80% RH.
2. Test cable length: 0m
3. Open and short calibration has been performed
4. D ≤ 0.1 for C, L measurements; Q ≤ 0.1 for R
measurements.
See the operation manual for additional details.