LCR-914/915/916 User Manual
25
(6) Resistance
Frequency
Range
Accuracy
100Hz
120Hz
200.00Ω
± (0.2% + 5d)
2.0000kΩ
± (0.2% + 2d)
20.000kΩ
200.00kΩ
2.0000MΩ
20.000MΩ
± (0.5% + 2d)
200.00MΩ
± (1.0% + 2d)
[2]
1kHz
20.000Ω
± (0.5% + 5d)
[2]
200.00Ω
± (0.2% + 2d)
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
20.000MΩ
± (2.0% + 2d)
200.00MΩ
± (5.0% + 2d)
[2]
10kHz
[1]
20.000Ω
± (0.5% + 5d)
[2]
200.00Ω
± (0.2% + 2d)
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
± (2.0% + 2d)
20.000MΩ
± (5.0% + 2d)
100kHz
[3]
20.000Ω
± (0.5% + 5d)
[2]
200.00Ω
± (0.2% + 2d)
2.0000kΩ
20.000kΩ
200.00kΩ
± (2.0% + 2d)
2.0000MΩ
± (5.0% + 2d)
[1] The 10kHz test frequency is only for the LCR-915/916.
[2] The measuring time is 2 seconds.
[3] The 100kHz test frequency is only for the LCR-916.
Input Protection: 30V
DC
or 30V
AC
rms
Minimum Resolution: 0.001Ω in the 20.000Ω range.
Measuring Time: 800ms
Note: If Q > 0.1, the accuracy should be multiplied by
2
1 Q