Figure 2 Access cover removal
Figure 3 Replace the fuses (continued)
1 Relay fuse (4x) 2 Output power fuse (2x) 3 Input power fuse (2x)
Prepare for storage or shipping
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Remove all the fluids and the power from the analyzer for long-term storage.
1. Push menu and then go to SETUP SYSTEM>DECOMISSIONING.
2. Select YES to stop the analyzer and start the decommissioning cycle.
3. Remove the reagents, the standards and the cleaning solution bottles for disposal in accordance
with local, state, and federal regulations.
4. Flush and then fill the bottles with deionized water.
5. Install the bottles.
6. Make sure that all of the liquid is drained from the colorimeter, sample holder and tubing.
English