Option Description
Jump to oldest Shows the first measurement.
Jump to Date & Time Shows the measurement recorded at the entered date and time.
3. To erase the data log, push MENU and select General Settings>Clear Data. Select Data Log,
then push Yes.
Show the event log
Show the event log to see the events that occurred.
1. Push MENU.
2. Select Diagnostics>Event Log.
The total number of events recorded shows followed by the total number of occurances for each
event type.
3. Select All Events or an event type. The time, date and description of each event shows.
4. To show any additional data for a selected event, select the event and push the RIGHT arrow.
5. To erase the event log, push MENU and select General Settings> Clear Data. Select Event Log,
then push Yes.
Show the alarm log
Show the alarm log to see the channel alarms that have occurred.
Note: Only the channel alarms configured by the user are recorded to the alarm log. Refer to Select and configure
the alarms on page 11 to select and configure the channel alarms that are recorded.
1. Push MENU.
2. Select Diagnostics>Alarm Log.
The total number of alarms recorded and the total number of occurances for each alarm type
show.
3. Select All Alarms or an alarm type. The alarm time and date and the alarm description for each
alarm shows.
4. To show any additional data for a selected alarm, select the event and push the RIGHT arrow.
5. To show the newest alarm, oldest alarm or the alarms that occurred on a specific date and time,
push Options, then select an option.
6. To erase the alarm log, push MENU and select General Settings> Clear Data. Select Alarm Log,
then push Yes.
Save logs and settings to a USB drive
N O T I C E
When the import option is used, all the user settings on the sampler are replaced with the selected user settings
on the USB flash drive. The data in the log files is erased.
Use the export option to:
Save a copy of the log files
5
to a USB flash drive.
Save a backup copy of the user settings (e.g., sampling program and hardware settings) to a USB
flash drive.
Use the import option to:
Replace the user settings on the sampler with a backup copy of the user settings.
Replace the user settings on the sampler with the user settings supplied by technical support or
from another sampler.
5
The log files are saved in FSDATA Desktop format.
English 21

Table of Contents

Other manuals for Hach AS950

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach AS950 and is the answer not in the manual?

Hach AS950 Specifications

General IconGeneral
Sample FrequencyProgrammable
TypePortable
Sampling MethodVacuum Pump
Temperature ControlRefrigerated
Bottle Configuration24 x 1L bottles
Power Supply12V DC
CommunicationRS-232
Sampler TypePortable or Stationary
Vertical Lift8.2 m
Enclosure RatingIP67
Number of SamplesUp to 999
Storage Capacity24 samples
Bottle Options1L