EasyManua.ls Logo

Hach DR/850 - CHLORINE, FREE, Ultra-high Range, DPD Method

Hach DR/850
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
101
CHLORINE, FREE, Ultra-high Range (0.0–10.0 mg/L Cl
2
) Method 10069
DPD Method USEPA accepted for reporting drinking water analyses
*
For testing higher levels of free chlorine (hypochlorous acid
and hypochlorite) in drinking water, cooling water,
and industrial process waters
Note: This product has not been evaluated to test for chlorine and chloramines in medical
applications in the United States.
* Procedure is equivalent to USEPA method 330.5 for wastewater and Standard Method 4500-C1-G for drinking
water.
1. Enter the user
program number for
Chlorine, UHR.
Press:
PRGM
The display will show:
PRGM?
Note: If the chlorine is
typically less than 2.0 mg/
L, use method 8021,
program number 9.
2. Press:
12 ENTER
The display will show
mg/L Cl
2
then: ZERO
3. Fill the 10-mL/1-cm
cell to the 5-mL line
with sample.
4. Place the cell into
the instrument. Cover
the sample cell tightly
with the instrument cap.
Note: Place the cell into
the cell holder as
illustrated. The sample cell
tab should be at the 6
o’clock position and
completely seated in the
cell holder slot.
mL
mL
5
10
Tab
Slot

Table of Contents

Other manuals for Hach DR/850

Related product manuals