Hach DR/850 - MOLYBDENUM, MOLYBDATE, Low Range Ternary Complex Method

Hach DR/850
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
219
Method 8169
MOLYBDENUM, MOLYBDATE, Low Range (0 to 3.00 mg/L)
Ternary Complex Method For boiler and cooling tower waters
1. Enter the stored
program number for
molybdate molybdenum.
Press:
PRGM
The display will show:
PRGM ?
Note: For most accurate
results, perform a Reagent
Blank Correction using
deionized water (see
Section 1).
2. Press: 47 ENTER
The display will show
mg/L, Mo6 and the
ZERO icon.
Note: For alternate forms
(MoO
4
), press the
CONC key.
3. Fill a 25-mL mixing
graduated cylinder with
20 mL of the sample.
Note: Filter turbid samples
using the labware listed
under Optional Apparatus.
4. Add the contents of
one Molybdenum 1
Reagent Powder Pillow
to the graduated
cylinder. Stopper. Invert
the graduated cylinder
several times to dissolve
the reagents.
5. Pour 10 mL of the
solution into a
sample cell.
6. Add 0.5 mL of
Molybdenum 2 Reagent
to the sample cell. Swirl
to mix. This is the
prepared sample.
Note: Molybdenum will
cause a green color to form.
7. Press:
TIMER ENTER
A two-minute reaction
period will begin.
8. Fill a second sample
cell with 10 mL of
solution from the
graduated cylinder
(the blank).