EasyManua.ls Logo

Hach DR/850 - CHLORINE, TOTAL, DPD Test N Tube Method

Hach DR/850
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
141
Method 10101
CHLORINE, TOTAL (0 to 5.00 mg/L) For water, wastewater and seawater
DPD Test ‘N Tube
Method
*
Note: This product has not been evaluated to test for chlorine and chloramines in medical
applications in the United States.
* Adapted from Standard Methods for the Examination of Water and Wastewater.
1. Enter the stored
program number for
Test ‘N Tube total
chlorine (Cl
2
).
Press:
PRGM
The display will show:
PRGM ?
Note: For most accurate
results, perform a Reagent
Blank Correction using
deionized water (see
Section 1).
2. Press: 10 ENTER
The display will show
mg/L, Cl2 and the
ZERO icon.
3. Insert the COD/TNT
Vial Adapter into the
cell holder by rotating
the adapter until it drops
into place. Then push
down to fully insert it.
Note: For increased
performance, a diffuser
band covers the light path
holes on the adapter. Do
not remove the diffuser
band.
4. Fill an empty Test ‘N
Tube vial with sample
(the blank).
Note: Fill to the top of the
Hach logo “oval” mark.
Note: Samples must be
analyzed immediately and
cannot be preserved for
later analysis.

Table of Contents

Other manuals for Hach DR/850

Related product manuals