EasyManua.ls Logo

Hach DR/850 - Page 211

Hach DR/850
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
211
d) Add two additional 0.1-mL aliquots of standard,
measuring the concentration after each addition. The iron
concentration should increase by 0.25 mg/L for each 0.1-
mL addition of standard.
e) If these increases do not occur, see Standard Additions in
Section 1 for more information.
Standard Additions Method (AccuVac Ampuls)
a) Use a graduated cylinder to measure 25.0 mL of sample
into each of three 50-mL beakers.
b) Snap the neck off an Iron Ampule Standard, 25 mg/L Fe.
c) Using a TenSette Pipet, add 0.1, 0.2 and 0.3 mL of
standard, respectively, to the 50-mL beakers. Swirl to mix.
d) Fill a TPTZ AccuVac Ampul from each beaker.
e) Measure the concentration of each ampul according to the
procedure. The iron concentration should increase by 0.1
mg/L for each 0.1 mL addition of standard.
f) If these increases do not occur, see Standard Additions in
Section 1 for more information.
Standard Solution Method
Prepare a 0.4 mg/L iron working solution as follows:
a) Using Class A glassware, pipet 1.00 mL of Iron Standard
Solution, 100 mg/L Fe, into a 250-mL volumetric flask.
b) Dilute to volume with deionized water. Stopper and invert
repeatedly to mix. Prepare this solution daily.
Method Performance
Precision
In a single laboratory using a standard solution of 1.00 mg/L Fe
and two representative lots of reagent with the instrument, a
single operator obtained a standard deviation of ±0.017 mg/L Fe.
In a single laboratory using a standard solution of 1.00 mg/L Fe
and one representative lot of AccuVac Ampuls with the
instrument, a single operator obtained a standard deviation of
±0.022 mg/L Fe.
IRON, TOTAL, continued

Table of Contents

Other manuals for Hach DR/850

Related product manuals