EasyManua.ls Logo

Hach DR/850 - Page 231

Hach DR/850
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
231
3. Analyze each spiked sample, following all steps of the
Monochloramine and Free Ammonia procedure. The
ammonia nitrogen concentration should increase 0.02 mg/L
for each 0.1 mL of standard added.
4. If these increases do not occur, see Standard Additions
(Section 1 of the DR/890 Procedures Manual) for more
information.
Standard Solution Method
Prepare a 0.20 mg/L ammonia nitrogen standard by diluting
2.00 mL of the Ammonia Nitrogen Standard Solution, 10 mg/L,
to 100 mL with dilution water. Or, using the TenSette Pipet,
prepare a 0.20 mg/L ammonia nitrogen standard by diluting
0.4 mL of a Ammonia Nitrogen Voluette Standard Solution,
50 mg/L as NH
3
–N, to 100 mL with dilution water. Analyze the
standard solution, following all steps of the Monochloramine and
Free Ammonia procedure.
Method Performance
Monochloramine Test
Precision
In a single laboratory, using a monochloramine standard solution
of 2.10 mg/L Cl
2
and representative lots of reagent, a single
operator obtained a standard deviation of ± 0.12 mg/L Cl
2
.
Estimated Detection Limit
The estimated detection limit for Method 10171 is 0.05 mg/L Cl
2
.
Free Ammonia Test
Precision
In a single laboratory using a solution containing 1.80 mg/L Cl
2
plus 0.20 mg/L ammonia nitrogen (NH
3
–N) and two
representative lots of reagent with the DR/890, a single operator
obtained a standard deviation of ± 0.01 mg/L N for seven
replicates.
Estimated Detection Limit
The estimated detection limit for program 46 is 0.02 mg/L N.
For more information on the estimated detection limit, see
Section 1 of the DR/850 or DR/890 Procedure Manual.
Nitrogen, Free Ammonia and Chloramine (Mono), continued

Table of Contents

Other manuals for Hach DR/850

Related product manuals