EasyManua.ls Logo

Hach DR/850 - Page 335

Hach DR/850
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
335
Estimated Detection Limit
The estimated detection limit for the programs #72, #73, and #74
is 0.02 mg/L O
3
. For more information on the estimated detection
limit, see Section 1.
Summary of Method
The reagent formulation adjusts the sample pH to 2.5 after the
ampul has filled. The indigo reagent reacts immediately and
quantitatively with ozone. The blue color of indigo is bleached in
proportion to the amount of ozone present in the sample. Other
reagents in the formulation prevent chlorine interference. No
transfer of sample is needed in the procedure. Therefore, ozone
loss due to sampling is eliminated.
REQUIRED REAGENTS
Quantity Required
Description Per Test Unit Cat. No.
Ozone AccuVac Ampuls
Select one or more based on range:
0-0.25 mg/L............................................................ 2 ampuls.................25/pkg..........25160-25
0-0.75 mg/L............................................................ 2 ampuls.................25/pkg..........25170-25
0-1.50 mg/L............................................................ 2 ampuls.................25/pkg..........25180-25
Water, deionized........................................................varies ........................ 4 L..............272-56
REQUIRED APPARATUS
Beaker, 50 mL...............................................................2 ...........................each ...........500-41H
OPTIONAL APPARATUS
AccuVac Snapper Kit ..................................................................................each ..........24052-00
AccuVac Ampule sampler...........................................................................each..........24051-00
For Technical Assistance, Price and Ordering
In the U.S.A.—Call 800-227-4224
Outside the U.S.A.—Contact the Hach office or distributor serving you.
OZONE, continued

Table of Contents

Other manuals for Hach DR/850