Do a test
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a
treatment process and/or chemical feed system for which there are
regulatory limits and monitoring requirements related to public health,
public safety, food or beverage manufacture or processing, it is the
responsibility of the user of this instrument to know and abide by any
applicable regulation and to have sufficient and appropriate
mechanisms in place for compliance with applicable regulations in
the event of malfunction of the instrument.
D A N G E R
Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in
accordance with local, regional and national regulations.
The generic steps to do a test follow.
To do a test for a specific parameter (e.g., chlorine), download the test
procedure from the manufacturer's website. Refer to Download a test
procedure on page 14.
1. Push to select the applicable measurement range (e.g., LR or
HR).
2. Prepare the blank. Refer to the test procedure.
3. Clean the sample cell with a no-lint cloth.
4. Insert the blank sample cell into the cell holder. Make sure to install
the blank sample cell in the correct and consistent orientation so
that the results are more repeatable and precise. Refer to Figure 5.
5. Install the instrument cap over the cell holder. Refer to Figure 6.
6. Push to set the instrument zero.
English 11

Other manuals for Hach DR300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach DR300 and is the answer not in the manual?

Hach DR300 Specifications

General IconGeneral
TypePortable Spectrophotometer
Operating Temperature0 to 50 °C
MethodPre-programmed methods
Light SourceLED
Wavelength Accuracy±2 nm
Resolution0.001 Abs
DisplayLCD
Measurement MethodAbsorbance
ParameterVarious
Photometric Range0 to 2.5 A
Dimensions230 x 93 x 280 mm (9.1 x 3.7 x 11.0 in)
Photometric Accuracy±0.005 Abs

Related product manuals