EasyManuals Logo

Hach DR300 User Manual

Hach DR300
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #3 background imageLoading...
Page #3 background image
Table of contents
Product components on page 6 Calibration on page 15
Install the batteries on page 7 Maintenance on page 20
User interface and navigation
on page 8
Troubleshooting on page 21
Do a test on page 11 Replacement parts and accessories
on page 24
Specifications
Specifications are subject to change without notice.
Specification Details
Dimensions (W x H x D) 6.9 x 15.7 x 3.4 cm (2.7 x 6.2 x 1.3 in.)
Enclosure IP67, waterproof at 1 m (3.3 ft) for 30 minutes when
battery compartment is closed and locked.
Light source Light emitting diode (LED)
Detector Silicon photodiode
Display LCD with backlight
Weight 0.25 kg (0.55 lb)
Power requirements 4 AAA batteries; approximate life of 5000 tests (use of
backlight decreases this number)
Rechargeable batteries are not recommended.
Operating environment 0 to 50 °C (32 to 122 °F), 0 to 90% relative humidity
non-condensing
Storage temperature –20 to 55 °C (–4 to 131 °F), 0 to 80% relative humidity
non-condensing
Wavelength Fixed wavelength ±2 nm, different for each model
Filter bandwidth 15 nm
Absorbance range 0 to 2.5 Abs
Sample cell 25 mm (10 mL) and 1 cm (10 mL)
Data storage Last 50 measurements
English 3

Other manuals for Hach DR300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach DR300 and is the answer not in the manual?

Hach DR300 Specifications

General IconGeneral
TypePortable Spectrophotometer
Operating Temperature0 to 50 °C
MethodPre-programmed methods
Light SourceLED
Wavelength Accuracy±2 nm
Resolution0.001 Abs
DisplayLCD
Measurement MethodAbsorbance
ParameterVarious
Photometric Range0 to 2.5 A
Dimensions230 x 93 x 280 mm (9.1 x 3.7 x 11.0 in)
Photometric Accuracy±0.005 Abs