Table 1 HQ Series meter models
Meter model Probe
connectors
pH/mV/ORP
1
probes
LDO/LBOD
2
probes
Conductivity
3
probes
ISE
4
probes
HQ1110 1
HQ1130 1
HQ1140 1
HQ2100 1
HQ2200 2
HQ4100 1
HQ4200 2
HQ4300 3
Section 3 Specifications
Specifications are subject to change without notice.
Specification Details
Dimensions (L x W x H) 22 x 9.7 x 6.3 cm (8.7 x 3.8 x 2.5 in.)
Weight HQ1110, HQ1130, HQ1140 and HQ2100: 519 g (18.3 oz); HQ2200: 541 g
(19.1 oz); HQ4100: 530 g (18.7 oz); HQ4200: 550 g (19.4 oz); HQ4300: 570 g
(20.1 oz)
Enclosure rating IP67 with battery compartment installed
Power requirements (internal) Rechargeable lithium-ion battery 18650 (18-mm diameter x 65-mm length,
cylindrical), 3.7 VDC, 3400 mAh; Battery Life: > 1 week with typical use
(10 readings/day, 5 days/week in Continuous or Push to read mode, or >
24 hours in Interval mode with 5 minute intervals and shutdown timer ≤
15 minutes)
Power requirements
(external)
Class II, USB power adapter: 100–240 VAC, 50/60 Hz input; 5 VDC at 2 A USB
power adapter output
Meter protection class IEC Class III (SELV (Separated/Safety Extra-Low Voltage) powered); USB power
adapter is IEC Class II (double-insulated)
Operating temperature 0 to 60 °C (32 to 140 °F)
Charging temperature 10 to 40 °C (50 to 104 °F)
Operating humidity 90% (non-condensing)
Storage temperature –20 to 60 °C (–4 to 140 °F) maximum 90% relative humidity (non-condensing)
Probe connector 5-pin M12 connector for Intellical probes
Micro-USB connector The micro-USB connector enables USB cable and USB power adapter
connectivity.
Data log (internal) HQ1000 Series: 5000 data points; HQ2000 Series: 10,000 data points;
HQ4000 Series: 100,000 data points
1
pH/mV/ORP probes include temperature
2
LDO/LBOD probes include temperature
3
Conductivity probes include salinity, TDS (total dissolved solids), resistivity, temperature
4
Ion-selective probes such as ammonia, nitrate, chloride, fluoride, sodium
4 English