Why sample properties affect stabilization in Hach ISENH4181 Laboratory Equipment?
- DDenise DixonAug 6, 2025
Sample properties can cause slow stabilization or inaccurate measurements in Hach Laboratory Equipment. Ensure the sample pH after adding ISA is less than 8.5, and use one ISA pillow per 25 mL of sample. Also, increase the sample temperature or keep the temperature consistent (within 2 °C (3.6 °F)) across different samples.