What to do if verification failed on Hach TU5200?
If verification fails, examine the verification standard. Then, calibrate the instrument. If verification is still unsuccessful, push Calibration>Options>Calibration Setup>Reset to Factory Calibration>OK.
What does 'Next calibration is due!' mean on Hach Laboratory Equipment?
Calibrate the instrument. Note that the calibration reminder is set to on.
What to do if there is a Calibration/Verification issue on Hach TU5200?
Examine the standards. Start the calibration or verification again. If calibration (or verification) is not successful, contact technical support.
What does 'Next verification is due!' mean on Hach TU5200?
Do a calibration verification. Note that the verification reminder is set to on.
What to do if the sample ID list is full on Hach TU5200?
Remove the sample IDs that are not used. Add a new sample ID.
What to do if setting the IP-address failed on Hach TU5200?
Contact the network administrator. If necessary, contact technical support.
How to fix incorrect date error on Hach TU5200?
Ensure that the date and time format is dd.mm.yyyy hh:mm.
Why is the standard value too high on my Hach TU5200?
A standard value that is too high may indicate that the wrong calibration standard is in the vial compartment. Ensure that the standard hasn't expired and put the correct calibration standard in the vial compartment.
What does it mean when the measured value is out of nominal value range on Hach TU5200?
This means the measured value is not within the expected range for the standard you are using. For example, the 10 NTU standard should measure between 5 to 15 NTU, and the < 0.1 NTU glass verification rod should measure between 0 to 0.1 NTU. Use the correct standard, ensure it hasn't expired, and measure it again.
| Brand | Hach |
|---|---|
| Model | TU5200 |
| Category | Laboratory Equipment |
| Language | English |
Crucial safety warnings and hazard information for operating the instrument.
Detailed information and warnings regarding the Class 1 laser classification.
Specific safety precautions related to the instrument's RFID module.
Recommended practices for placing and setting up the instrument.
Settings and parameter adjustments for instrument operation.
Adjusting parameters like reading mode, units, and data logging.
Procedures for taking and recording sample measurements.
The process of initiating and completing a sample measurement.
Comparing measurements from different sources or instruments.
Step-by-step guide to perform the instrument calibration process.
Executing the steps to verify the instrument's calibration.