EasyManuals Logo

Hach TU5200 User Manual

Hach TU5200
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
Option Description
Verify after Cal. Sets the instrument to start a verification immediately after the instrument is calibrated.
When set to on, the verification standard is measured immediately after a calibration is
done. Default: ON.
Acceptance
Range
Sets the maximum difference permitted between the recorded value of the verification
standard and the measured value of the verification standard during verification. Unit
Sets the acceptance range for verification to a percentage (1 to 99%) or an NTU value
(0.015 to 100.00 NTU). Options: % or NTU (or mNTU).
Verification
Reminder
Sets the time interval between calibration verifications. The display will show a reminder
when a verification is due. Options: OFF(default), 1 day, 7 days, 30 days or 90 days.
When a verification is done, the verification time is set to zero.
Do a calibration verification
Do a calibration verification to make sure that the instrument is still calibrated as necessary.
1. Log in to the instrument as follows:
Put an operator RFID tag in front of the RFID module or
Push Login. Select the applicable operator ID, then push Select.
2. Push Verification.
3. Clean the verification standard vial with a no-lint cloth to remove contamination.
Keep contamination off the glass. Refer to Prevent vial contamination on page 18.
4. Dry the external surfaces of the vial with a no-lint cloth. Make sure to dry the bottom of the vial.
5. Carefully invert the vial three times. Refer to Figure 6 on page 27.
6. Put the vial in the vial compartment.
7. Push Start.
8. Confirm the value shown or enter the exact turbidity value of the standard. Push Read.
9. When the measurement is complete, remove the vial.
The verification is recorded to the verification log.
10. To show the verification log, push Options>Verification Log. Refer to Show the recorded data
on page 24 for more options.
11. To send the verification data to external devices that are connected to the instrument, push
Options>Send Data. Refer to Show the recorded data on page 24 for more options.
Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
28 English

Other manuals for Hach TU5200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach TU5200 and is the answer not in the manual?

Hach TU5200 Specifications

General IconGeneral
BrandHach
ModelTU5200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals