Measure the sample
1. Push Read if a measurement does not start automatically when the lid is closed.
2. When the measurement is complete, push Options>Store to record the measurement to the
reading log as necessary.
Note: If the Auto Save setting is set to on, "Data Stored" shows on the display and the measurement is
automatically recorded to the reading log.
3. To show the recorded measurements, push Options>Reading Log. Refer to Show the recorded
data on page 24 for more options.
4. To send the measurement data to external devices that are connected to the instrument, push
Options>Send Data. Refer to Show the recorded data on page 24 for more options.
Note: If the Auto Send settings is set to on, the measurement data is automatically sent to the external
device(s) that is connected to the instrument.
Compare process and laboratory measurements
Compare process and laboratory measurements with RFID or Link2SC. Make sure that the process
and lab instrument are calibrated with the same number of calibration points and with the same
standards. Make sure that the calibrations are not expired.
Collect a grab sample
Collect a 100-mL sample (minimum) from the sample outlet tubing of the process instrument. Collect
the sample in a clean glass bottle with a tight-fitting cap. Do not collect samples directly into a sample
vial.
1. Rinse the glass bottle a minimum of three times with water from the sample outlet tubing of the
process instrument. Let the bottle overflow with the sample.
2. Collect a 100-mL sample (minimum) in the glass bottle from the sample outlet tubing of the
process instrument.
3. Put the cap on the sample bottle.
4. Analyze the grab sample immediately with the laboratory instrument to prevent settling, bacteria
growth and temperature changes.
Compare measurements with RFID
When the process instrument and laboratory instrument have the optional RFID module, compare
process and laboratory measurements with RFID.
Items to collect:
TU5300 sc or TU5400 sc with the optional RFID module
TU5200 with the optional RFID module
TU5200 sample vials
English
21

Other manuals for Hach TU5200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach TU5200 and is the answer not in the manual?

Hach TU5200 Specifications

General IconGeneral
BrandHach
ModelTU5200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals