Catalog Number 8990
Sigma 900 MAX Refrigerated Sampler
INSTRUMENT MANUAL
07/03 4ed

Table of Contents

Questions and Answers

  • S
    Sonia HodgeSep 8, 2025
    Why does Hach Laboratory Equipment display 'Tried to enable Storm Water with a single bottle'?
    • O
      Olivia RhodesSep 8, 2025
      This message appears because Stormwater mode is not allowed when the sampler is configured with only one bottle. Use multiple bottles to enable Stormwater mode.
  • S
    Stephanie HolmesAug 25, 2025
    What does 'Sensor output voltage too low—cannot calibrate' mean on Hach 900 MAX and how to fix it?
    • M
      Manuel WilsonAug 25, 2025
      The message 'Sensor output voltage too low—cannot calibrate' indicates that the D.O. calibration yielded an invalid result. Try another probe.
  • J
    jennifernolanAug 25, 2025
    What does 'Calibration requires a minimum submersion of six inches' mean on Hach 900 MAX?
    • T
      Tabitha FerrellAug 26, 2025
      This message relates to level sensor calibration of the bubbler or submerged sensor. Ensure a minimum submersion of six inches during calibration.
  • S
    Sherry MayerSep 18, 2025
    What does 'Hours must be: 01—12' mean on Hach 900 MAX Laboratory Equipment and how to fix it?
    • J
      Jason MillerSep 18, 2025
      This message indicates that an invalid number was entered when setting a time, in 12-hour format. Enter a valid hour in 12-hour format.
  • A
    Anthony SmithSep 8, 2025
    What does 'The program setup needs flow to be logged at an interval < or = the flow time limit' mean on Hach 900 MAX and how to resolve it?
    • Q
      qwalkerSep 9, 2025
      This message indicates that an alarm, setpoint, or stormwater function uses flow rate-of-change as a trip point, and the logging interval is not adequate. Adjust the logging interval to meet the requirements.
  • R
    Richard DavisSep 13, 2025
    Why does Hach 900 MAX Laboratory Equipment display 'Upset sampling is not allowed when the sampler is configured with one bottle'?
    • P
      Phillip HarmonSep 13, 2025
      This message appears when you try to enable Upset Sampling with a single bottle. Use multiple bottles to enable Upset Sampling.
  • R
    Ronald MillerSep 3, 2025
    Why does Hach 900 MAX display 'No Change in Level Readings or Inaccurate Level Readings'?
    • T
      Tracy BrownSep 3, 2025
      The message 'No Change in Level Readings or Inaccurate Level Readings' is displayed because of calibration issues. Re-calibrate the unit.
  • M
    Mr. Robert HarrisonSep 4, 2025
    What does 'Function disabled while a program is running. Use RUN/STOP key' mean on Hach 900 MAX Laboratory Equipment?
    • J
      jason95Sep 4, 2025
      The message 'Function disabled while a program is running. Use RUN/STOP key' means you tried to use manual menu functions or the distributor test while the program is running. Use the RUN/STOP key to stop the program before retrying.
  • S
    Sara SmithSep 23, 2025
    What does 'No year entered' mean on Hach 900 MAX and how to fix it?
    • P
      Paula TrujilloSep 23, 2025
      This message indicates that an invalid number was entered when setting a date. Enter a valid year.
  • E
    Elizabeth WhiteAug 29, 2025
    Why does Hach 900 MAX show 'U-Sonic Failure' and how to fix it?
    • S
      Shannon MarshAug 29, 2025
      A 'U-Sonic Failure' message indicates that the transducer is either not plugged in, the cable is damaged, or the transducer thermal sensor is damaged. Verify that the transducer is correctly plugged in and inspect the sensor for any signs of damage.

Summary

Safety Precautions

Use of Hazard Information

Explains hazard signal words (Danger, Caution, Note) for understanding warnings.

Precautionary Labels

Advises reading and observing all labels attached to the instrument for safety.

Hazardous Locations

States the instrument is not approved for hazardous locations as defined by the National Electrical Code.

Specifications

General and Hardware Specifications

Covers dimensions, cabinet, thermal system, power, and control panel.

Sampling and Communication Features

Details sampling modes, intervals, communication, and data logging.

Optional Device and Sensor Specifications

Lists specifications for optional probes (pH, DO, ORP, Conductivity) and sensors.

Installation

Unpacking and Site Selection

Guides on unpacking and choosing an optimal installation site.

Basic Programming Setup

Initial Power-Up and Menu Navigation

Guides on initial power-up, main menu, and basic programming steps.

Core Program Configuration Steps

Details setup steps for bottles, intake tubing, program lock, and sample collection.

Sensor Setup and Calibration

Maintenance Procedures