Catalog Number 8990
Sigma 900 MAX Refrigerated Sampler
INSTRUMENT MANUAL
07/03 4ed

Table of Contents

Questions and Answers

  • R
    Robert RamirezSep 2, 2025
    What does 'Sampling halted during rinse—no liquid detected' mean for Hach Laboratory Equipment and what to do?
    • T
      Todd PowellSep 2, 2025
      The message 'Sampling halted during rinse—no liquid detected' means that a sample attempt was stopped because no liquid was detected during rinse. Check the sampling setup.
  • J
    Jeffrey BakerSep 2, 2025
    What does 'Sampling halted due to fluid presence after rinse purge' mean for Hach Laboratory Equipment and what to do?
    • S
      Steven BridgesSep 2, 2025
      The message 'Sampling halted due to fluid presence after rinse purge' means that a sample attempt was stopped because fluid was detected after rinse purge. Investigate the tubing setup.
  • R
    Ronald MillerSep 3, 2025
    Why does Hach 900 MAX display 'No Change in Level Readings or Inaccurate Level Readings'?
    • T
      Tracy BrownSep 3, 2025
      The message 'No Change in Level Readings or Inaccurate Level Readings' is displayed because of calibration issues. Re-calibrate the unit.
  • M
    Mr. Robert HarrisonSep 4, 2025
    What does 'Function disabled while a program is running. Use RUN/STOP key' mean on Hach 900 MAX Laboratory Equipment?
    • J
      jason95Sep 4, 2025
      The message 'Function disabled while a program is running. Use RUN/STOP key' means you tried to use manual menu functions or the distributor test while the program is running. Use the RUN/STOP key to stop the program before retrying.
  • D
    donald30Sep 4, 2025
    Why does Hach Laboratory Equipment say 'Logged data will be lost if you choose to continue?'
    • J
      Joshua PetersonSep 4, 2025
      This message appears when making a program change while a program is halted. Choose 'No' to avoid losing logged data.
  • S
    Stephanie KennedySep 5, 2025
    Why does Hach Laboratory Equipment display 'Error: Distributor arm is not functioning properly!'?
    • L
      Lauren RogersSep 5, 2025
      The error 'Distributor arm is not functioning properly!' indicates that the distributor test failed. This could be due to an arm obstruction, a cable that is not properly installed, or a defect in the distributor assembly. Check for obstructions, ensure the cable is correctly installed, and inspect the distributor assembly for defects.
  • N
    nancymillerSep 5, 2025
    Why does Hach Laboratory Equipment display 'The program cannot start: timed rinse required'?
    • R
      Ryan WilsonSep 6, 2025
      This message appears if the liquid sensor is disabled and timed calibration has not been performed for the rinse cycle. Enable the liquid sensor and perform timed calibration.
  • S
    Steven MorrisSep 6, 2025
    Why does Hach 900 MAX Laboratory Equipment display 'The program cannot start: volume calibration required'?
    • H
      Hannah ReidSep 6, 2025
      This message appears if the liquid sensor is disabled and timed calibration has not been performed for the main sample, the first flush, or the upset sample volume. Enable the liquid sensor and perform volume calibration.
  • D
    Diana Rodriguez DVMSep 7, 2025
    What to do if Hach Laboratory Equipment displays 'The program cannot start: the primary device is not properly set up'?
    • A
      april82Sep 7, 2025
      This message appears if the operator goes partially through the flow meter setup screens but stops before entering all necessary parameters. Complete all parameters in the flow meter setup.

Summary

Safety Precautions

Use of Hazard Information

Explains hazard signal words (Danger, Caution, Note) for understanding warnings.

Precautionary Labels

Advises reading and observing all labels attached to the instrument for safety.

Hazardous Locations

States the instrument is not approved for hazardous locations as defined by the National Electrical Code.

Specifications

General and Hardware Specifications

Covers dimensions, cabinet, thermal system, power, and control panel.

Sampling and Communication Features

Details sampling modes, intervals, communication, and data logging.

Optional Device and Sensor Specifications

Lists specifications for optional probes (pH, DO, ORP, Conductivity) and sensors.

Installation

Unpacking and Site Selection

Guides on unpacking and choosing an optimal installation site.

Basic Programming Setup

Initial Power-Up and Menu Navigation

Guides on initial power-up, main menu, and basic programming steps.

Core Program Configuration Steps

Details setup steps for bottles, intake tubing, program lock, and sample collection.

Sensor Setup and Calibration

Maintenance Procedures

Hach 900 MAX Specifications

General IconGeneral
BrandHach
Model900 MAX
CategoryLaboratory Equipment