Sigma 900 MAX Refrigerated Sampler
INSTRUMENT MANUAL
07/03 4ed

Table of Contents

Questions and Answers

  • J
    jennifer24Aug 27, 2025
    What causes 'Low Slate Memory' on Hach 900 MAX Laboratory Equipment and how to fix it?
    • K
      Kathryn PowellAug 27, 2025
      When free slate memory is less than 20%, the 'Low Slate Memory' error appears, indicating that RAM memory is almost full and will stop recording soon. To resolve this, you can download data from the unit, halt and restart the program. Another option is to download the data, halt the program, and change the data to wrap mode.
  • J
    Joel RichardAug 27, 2025
    How to resolve 'Incorrect password! Access denied' on Hach 900 MAX Laboratory Equipment?
    • J
      Jessica YoungAug 27, 2025
      If you see 'Incorrect password! Access denied,' it means the wrong password was entered. The correct password is 9000. Enter the correct password to proceed.
  • J
    Jennifer YatesAug 28, 2025
    What causes 'Jammed Distributor' error in Hach 900 MAX Laboratory Equipment and how to fix it?
    • J
      jessica63Aug 28, 2025
      If the distributor arm has jammed on an obstruction, the 'Jammed Distributor' error will appear. Remove the obstruction to resolve this issue.
  • S
    Shane SmithAug 28, 2025
    What to do if Hach 900 MAX Laboratory Equipment shows 'Memory Battery' error?
    • J
      Jared WilsonAug 29, 2025
      If the internal memory battery voltage is too low, the 'Memory Battery' error appears. To fix this, change the memory batteries.
  • E
    Elizabeth WhiteAug 29, 2025
    Why does Hach 900 MAX show 'U-Sonic Failure' and how to fix it?
    • S
      Shannon MarshAug 29, 2025
      A 'U-Sonic Failure' message indicates that the transducer is either not plugged in, the cable is damaged, or the transducer thermal sensor is damaged. Verify that the transducer is correctly plugged in and inspect the sensor for any signs of damage.
  • V
    Victoria SmithAug 30, 2025
    What does 'Modem Failure' mean on Hach Laboratory Equipment and how can I resolve it?
    • D
      dthomasAug 30, 2025
      A 'Modem Failure' indicates that the device is unable to initialize the modem PCB. If the modem is enabled, the problem may be on the modem board, and the board may need replacement.
  • P
    Paul JacksonAug 30, 2025
    What does 'U-Sonic Echo Loss' mean on Hach 900 MAX and how can I resolve it?
    • L
      Lynn PerezAug 30, 2025
      The 'U-Sonic Echo Loss' error means that the echo has been temporarily deflected by a change in site conditions, such as floating debris, foam in the channel, or wind. Check for excessive foam and remove any debris or build-up.
  • M
    Michael AustinAug 25, 2025
    What does 'pH calibration failed—gain and/or offset out of range. Try again' mean on Hach Laboratory Equipment and how to fix it?
    • M
      Margaret LeeAug 25, 2025
      The message 'pH calibration failed—gain and/or offset out of range. Try again' indicates that the pH calibration calculations yielded an invalid number. Try using fresh buffers or a new probe.
  • C
    Charles HawkinsSep 1, 2025
    What does 'Sampling halted because of key press' mean for Hach 900 MAX Laboratory Equipment and what to do?
    • N
      nicholaswilsonSep 1, 2025
      The message 'Sampling halted because of key press' means that a sample attempt was stopped because a key was pressed at the end of sampling. Avoid pressing keys during sampling.
  • A
    Anthony RichardsonSep 1, 2025
    What does 'Sampling halted due to fluid presence after pre-purge' mean for Hach 900 MAX and what to do?
    • B
      Brandon BrownSep 1, 2025
      The message 'Sampling halted due to fluid presence after pre-purge' means that a sample attempt was stopped because fluid was detected after pre-purge. Investigate the tubing setup or the presence of fluid.

Summary

Safety Precautions

Use of Hazard Information

Explains hazard signal words (Danger, Caution, Note) for understanding warnings.

Precautionary Labels

Advises reading and observing all labels attached to the instrument for safety.

Hazardous Locations

States the instrument is not approved for hazardous locations as defined by the National Electrical Code.

Specifications

General and Hardware Specifications

Covers dimensions, cabinet, thermal system, power, and control panel.

Sampling and Communication Features

Details sampling modes, intervals, communication, and data logging.

Optional Device and Sensor Specifications

Lists specifications for optional probes (pH, DO, ORP, Conductivity) and sensors.

Installation

Unpacking and Site Selection

Guides on unpacking and choosing an optimal installation site.

Basic Programming Setup

Initial Power-Up and Menu Navigation

Guides on initial power-up, main menu, and basic programming steps.

Core Program Configuration Steps

Details setup steps for bottles, intake tubing, program lock, and sample collection.

Sensor Setup and Calibration

Maintenance Procedures

Hach 900 MAX Specifications

General IconGeneral
BrandHach
Model900 MAX
CategoryLaboratory Equipment
LanguageEnglish