Sigma 900 MAX Refrigerated Sampler
INSTRUMENT MANUAL
07/03 4ed

Table of Contents

Questions and Answers

  • D
    Daisy BarronSep 22, 2025
    What does 'The value must be: xxx—yyy (xxx is minimum, yyy is maximum)' mean on Hach Laboratory Equipment and how to fix it?
    • D
      Douglas WrightSep 22, 2025
      This message indicates that the value entered was out of the legal range when changing a numeric parameter. Enter a value within the specified range.
  • S
    Stephanie EllisSep 23, 2025
    What does 'Logging intervals must be: 1,2,3,5,6,10,12,15,20,30,60' mean on Hach 900 MAX and how to fix it?
    • R
      Robert PhelpsSep 23, 2025
      This message indicates that an invalid logging interval was entered. Enter a valid logging interval.
  • J
    Jasmine WalshSep 23, 2025
    What does 'No changes allowed while a pgm is running. Use run/stop key' mean on Hach 900 MAX Laboratory Equipment?
    • J
      Justin RobinsonSep 23, 2025
      This message indicates that an inaccessible selection was made on a menu screen while a program is running. Stop the program before proceeding, using the RUN/STOP key.
  • J
    Joyce WhiteSep 23, 2025
    What does 'No retries are possible when the liquid sensor is disabled' mean on Hach 900 MAX Laboratory Equipment and what to do?
    • B
      Brett NelsonSep 23, 2025
      This message appears if Sample Retries is selected from the Modify Selected Items scrolling list, and the liquid sensor is disabled. Enable the liquid sensor to use retries.
  • S
    Sara SmithSep 23, 2025
    What does 'No year entered' mean on Hach 900 MAX and how to fix it?
    • P
      Paula TrujilloSep 23, 2025
      This message indicates that an invalid number was entered when setting a date. Enter a valid year.
  • J
    Jason BurgessAug 31, 2025
    What does 'Sampling halted because the bottle is full' mean for Hach 900 MAX and what to do?
    • H
      Holly ConnerAug 31, 2025
      The message 'Sampling halted because the bottle is full' means that a sample attempt was stopped because the bottle is full. Replace or empty the bottle before sampling again.
  • S
    Stephanie HolmesAug 25, 2025
    What does 'Sensor output voltage too low—cannot calibrate' mean on Hach 900 MAX and how to fix it?
    • M
      Manuel WilsonAug 25, 2025
      The message 'Sensor output voltage too low—cannot calibrate' indicates that the D.O. calibration yielded an invalid result. Try another probe.
  • J
    jennifernolanAug 25, 2025
    What does 'Calibration requires a minimum submersion of six inches' mean on Hach 900 MAX?
    • T
      Tabitha FerrellAug 26, 2025
      This message relates to level sensor calibration of the bubbler or submerged sensor. Ensure a minimum submersion of six inches during calibration.
  • C
    Christy RussellAug 26, 2025
    How to fix 'Low Main Battery' on Hach 900 MAX?
    • L
      Lisa JonesAug 26, 2025
      If the power supply voltage is less than 11 V dc, the 'Low Main Battery' error appears. To solve this, change the batteries.
  • D
    Daniel FinleyAug 27, 2025
    What to do if Hach 900 MAX Laboratory Equipment shows 'Slate Memory Full'?
    • J
      Jacob BarajasAug 27, 2025
      If your unit is in slate memory mode and can't log any more data because there is no more slate memory, download the data and restart the program. Alternatively, download the data and then change the memory mode.

Summary

Safety Precautions

Use of Hazard Information

Explains hazard signal words (Danger, Caution, Note) for understanding warnings.

Precautionary Labels

Advises reading and observing all labels attached to the instrument for safety.

Hazardous Locations

States the instrument is not approved for hazardous locations as defined by the National Electrical Code.

Specifications

General and Hardware Specifications

Covers dimensions, cabinet, thermal system, power, and control panel.

Sampling and Communication Features

Details sampling modes, intervals, communication, and data logging.

Optional Device and Sensor Specifications

Lists specifications for optional probes (pH, DO, ORP, Conductivity) and sensors.

Installation

Unpacking and Site Selection

Guides on unpacking and choosing an optimal installation site.

Basic Programming Setup

Initial Power-Up and Menu Navigation

Guides on initial power-up, main menu, and basic programming steps.

Core Program Configuration Steps

Details setup steps for bottles, intake tubing, program lock, and sample collection.

Sensor Setup and Calibration

Maintenance Procedures

Hach 900 MAX Specifications

General IconGeneral
BrandHach
Model900 MAX
CategoryLaboratory Equipment
LanguageEnglish