Questions and Answers

  • P
    Paul JacksonAug 30, 2025
    What does 'U-Sonic Echo Loss' mean on Hach 900 MAX and how can I resolve it?
    • L
      Lynn PerezAug 30, 2025
      The 'U-Sonic Echo Loss' error means that the echo has been temporarily deflected by a change in site conditions, such as floating debris, foam in the channel, or wind. Check for excessive foam and remove any debris or build-up.
  • S
    Sonia HodgeSep 8, 2025
    Why does Hach Laboratory Equipment display 'Tried to enable Storm Water with a single bottle'?
    • O
      Olivia RhodesSep 8, 2025
      This message appears because Stormwater mode is not allowed when the sampler is configured with only one bottle. Use multiple bottles to enable Stormwater mode.
  • E
    Elizabeth WhiteAug 29, 2025
    Why does Hach 900 MAX show 'U-Sonic Failure' and how to fix it?
    • S
      Shannon MarshAug 29, 2025
      A 'U-Sonic Failure' message indicates that the transducer is either not plugged in, the cable is damaged, or the transducer thermal sensor is damaged. Verify that the transducer is correctly plugged in and inspect the sensor for any signs of damage.
  • D
    Donald DanielsSep 22, 2025
    What does 'Invalid floating point format, only one decimal point allowed' mean on Hach 900 MAX Laboratory Equipment and how to correct it?
    • A
      Autumn OlsonSep 22, 2025
      This message indicates that a number was entered with two or more decimal points. Ensure the number contains only one decimal point.
  • T
    Thomas GilmoreSep 9, 2025
    What does 'The program setup needs rain to be logged at an interval < or = the rainfall time limit' mean on Hach Laboratory Equipment and how to resolve it?
    • F
      freytraceySep 9, 2025
      This message indicates that an alarm, setpoint, or stormwater function uses rainfall as a trip point, and the logging interval is not adequate. Adjust the logging interval to meet the requirements.
  • R
    Richard DavisSep 13, 2025
    Why does Hach 900 MAX Laboratory Equipment display 'Upset sampling is not allowed when the sampler is configured with one bottle'?
    • P
      Phillip HarmonSep 13, 2025
      This message appears when you try to enable Upset Sampling with a single bottle. Use multiple bottles to enable Upset Sampling.
  • S
    Sherry MayerSep 18, 2025
    What does 'Hours must be: 01—12' mean on Hach 900 MAX Laboratory Equipment and how to fix it?
    • J
      Jason MillerSep 18, 2025
      This message indicates that an invalid number was entered when setting a time, in 12-hour format. Enter a valid hour in 12-hour format.
  • D
    Daniel FinleyAug 27, 2025
    What to do if Hach 900 MAX Laboratory Equipment shows 'Slate Memory Full'?
    • J
      Jacob BarajasAug 27, 2025
      If your unit is in slate memory mode and can't log any more data because there is no more slate memory, download the data and restart the program. Alternatively, download the data and then change the memory mode.
  • C
    Christy RussellAug 26, 2025
    How to fix 'Low Main Battery' on Hach 900 MAX?
    • L
      Lisa JonesAug 26, 2025
      If the power supply voltage is less than 11 V dc, the 'Low Main Battery' error appears. To solve this, change the batteries.
  • J
    jennifernolanAug 25, 2025
    What does 'Calibration requires a minimum submersion of six inches' mean on Hach 900 MAX?
    • T
      Tabitha FerrellAug 26, 2025
      This message relates to level sensor calibration of the bubbler or submerged sensor. Ensure a minimum submersion of six inches during calibration.

Summary

Safety Precautions

Use of Hazard Information

Explains hazard signal words (Danger, Caution, Note) for understanding warnings.

Precautionary Labels

Advises reading and observing all labels attached to the instrument for safety.

Hazardous Locations

States the instrument is not approved for hazardous locations as defined by the National Electrical Code.

Specifications

General and Hardware Specifications

Covers dimensions, cabinet, thermal system, power, and control panel.

Sampling and Communication Features

Details sampling modes, intervals, communication, and data logging.

Optional Device and Sensor Specifications

Lists specifications for optional probes (pH, DO, ORP, Conductivity) and sensors.

Installation

Unpacking and Site Selection

Guides on unpacking and choosing an optimal installation site.

Basic Programming Setup

Initial Power-Up and Menu Navigation

Guides on initial power-up, main menu, and basic programming steps.

Core Program Configuration Steps

Details setup steps for bottles, intake tubing, program lock, and sample collection.

Sensor Setup and Calibration

Maintenance Procedures